Republished by Newlife - Radbar. All rights belong to the original publisher; see Source below.
A clinical trial evaluating the radiation protection requirements of a novel specimen positron emission tomography and computed tomography scanner has shown that occupational radiation doses to operating theatre personnel remain manageable and compliant with national regulatory limits during cancer resections. The study, conducted at the first hospital centre in the United Kingdom to deploy the mobile AURA10 specimen scanner manufactured by Xeos in Ghent and distributed by Southern Scientific Limited, tracked occupational radiation exposure across head and neck oncological resections as well as robotic radical prostatectomies. Investigated as part of the locally sponsored eXcision trial, registered under trial identifier NCT06676943, the platform is designed to provide immediate, high-resolution three-dimensional radiological margin assessment directly within operating suites. By introducing tumour-avid positron-emitting radiopharmaceuticals into patients before tissue excision, surgical teams aim to perform precise resections that remove the entire malignant lesion while sparing adjacent healthy tissue, thereby helping avoid secondary adjuvant chemoradiotherapy. However, administering radioactive tracers within surgical theatres introduces distinct health physics considerations, requiring rigorous quantification of occupational exposures, spatial dose rates, radioactive contamination pathways, and waste containment protocols across multidisciplinary teams.
The physical design and operational mechanics of the specimen scanner dictate both the workflow and the radiation environment within the surgical suite. The imaging unit combines an integrated ring of twenty-two lutetium-yttrium oxyorthosilicate crystal detectors with a fifty-kilovolt x-ray tube and an associated imaging plate positioned directly beneath. Once an excised surgical specimen is removed from the operative field, the surgeon places the tissue into a single-use dish measuring ten centimetres in diameter. This specimen container locks onto an automated stage that lowers into the x-ray chamber where it rotates to acquire high-resolution computed tomography projections. The container then elevates into the surrounding ring of crystal scintillation detectors to record positron annihilation photon emissions, following a standard acquisition protocol lasting ten minutes while the volumetric computed tomography datasets undergo mathematical reconstruction. Co-registered imaging is presented immediately on an integrated theatre display screen, allowing the surgical team to examine the specimen in three orthogonal planes, inspect maximum intensity projection renderings, or review isolated computed tomography and positron emission datasets. Radiation leakage surveys conducted by the manufacturer and independently replicated by institutional radiation physics personnel using a Fluke Biomedical 451P survey meter confirmed that no detectable ionizing leakage escaped the unit when the internal x-ray tube was energized.
Establishing the safety profile required precise radioactive drug administration schedules tailored to the physiological and surgical demands of each oncology subspecialty. In head and neck surgical procedures, patients received an intravenous injection of fluorine-18 fluorodeoxyglucose at an activity concentration of 0.8 megabecquerels per kilogram of body mass, which resulted in a mean administered activity of 56 megabecquerels across the cohort, ranging from a minimum of 36 megabecquerels to a maximum of 85 megabecquerels. For urology procedures, patients received a nominal fixed activity of 50 megabecquerels of gallium-68 prostate-specific membrane antigen, yielding a measured mean administered activity of 49 megabecquerels, spanning from 41 to 56 megabecquerels. Radiotracers were injected inside the operating theatre between one and two hours prior to surgical resection.
Spatial configurations within the operating theatre dictated the baseline distances and relative geometry between staff and the radioactive patient. In head and neck cases, one to two lead surgeons and designated assistant surgeons operated in close proximity to the patient, maintaining an approximate working distance of 0.1 metres from the patient for up to four hours following radiopharmaceutical injection. The lead scrub nurse stood at an estimated distance of one metre from the operating table throughout the surgical intervention, while the anaesthetic operating department practitioner and the primary anaesthetist occupied positions approximately 1.5 metres and 2.0 metres away, respectively. In contrast, radical prostatectomies were conducted using a surgical robotic platform where the console surgeon operated from an unshielded workstation located approximately three metres from the patient. An assistant surgeon stood at the patient bedside at an approximate distance of 0.5 metres throughout the intervention, accompanied by a scrub nurse at one metre and anaesthesia personnel positioned approximately 1.5 metres from the patient. The physical decay of the respective positron-emitting isotopes, combined with distance and biological excretion pathways modelled using publication data from the International Commission on Radiological Protection, established the mathematical foundation for prior statutory risk assessments.
To capture accurate occupational exposures across multiple surgical shifts over a ten-month clinical evaluation period, investigators deployed an extensive suite of personal dosimetry technologies. Whole-body deep dose equivalents were monitored using Landauer Luxel-plus optically stimulated luminescence dosimeters worn on the torso beneath standard theatre gowns, possessing a detection range extending from 0.01 to 10,000 millisieverts. Extremity doses to the fingers were recorded with Landauer Saturn thermoluminescent dosimeter rings placed on the index finger of the non-dominant hand with the sensing element oriented towards the palmar aspect, operating across an energy range above fifteen kiloelectronvolts. Ocular radiation exposures were monitored using Landauer Vision thermoluminescent dosimeters mounted on eyeglasses or protective theatre caps near the brow. Additionally, personnel were equipped with Tracerco personal electronic dosimeters worn on the abdomen at the whole-body dosimeter height to deliver instantaneous operational readings at the conclusion of each operative procedure. Dosimetric tracking spanned eighteen head and neck operations and thirty-eight urological operations, partitioned across three consecutive badge wear periods. Although dosimeter allocation followed structured departmental sign-in rosters, occasional shift disruptions and badge mix-ups occurred, requiring the exclusion of mismatched dosimeters from final analytical calculations.
Dosimetric results demonstrated that radiation exposure was strongly correlated with anatomical proximity to the patient, with lead surgeons and scrub nurses recording the highest occupational doses. In head and neck surgeries, primary operating surgeons received a measured mean whole-body dose of 39 microsieverts per procedure, accompanied by a mean finger extremity dose of 88 microsieverts per procedure. The assisting lead scrub nurse recorded a mean whole-body dose of 12 microsieverts and an extremity dose of 48 microsieverts per head and neck case. Eye doses for head and neck surgeons monitored with thermoluminescent dosimeters remained below recording threshold values across all evaluated procedures. In urology cases, where the console surgeon remained three metres away, the assistant surgeon positioned beside the patient recorded a mean whole-body dose of 7.2 microsieverts and a mean finger dose of 16 microsieverts per resection. The urology scrub nurse received a mean whole-body dose of 2.5 microsieverts per case, with finger doses remaining entirely below baseline detection thresholds. All other operating theatre personnel, including anaesthetists and operating department practitioners, received whole-body doses that fell well below statutory public dose limits.
Specialized physical engineering controls and dedicated mobile shielding stations were fabricated to mitigate radiation fields and handle contaminated surgical items during intraoperative care. Southern Scientific supplied bespoke stainless-steel equipment trolleys manufactured by Bright Technologies, which incorporated six millimetres of lead shielding. Because six millimetres corresponds precisely to the half-value layer thickness for 511-kiloelectronvolt positron annihilation gamma photons, the shielding provided substantial attenuation while preserving physical mobility within the operative space. A specialized catheter trolley housed indwelling urinary drainage bags for head and neck patients; surface dose rate monitoring demonstrated that the lead barrier achieved an average fourfold reduction in emitted radiation fields, attenuating a mean catheter bag surface dose rate of 11 microsieverts per hour down to low ambient levels. A secondary mobile trolley housed lead-shielded receptacles for contaminated surgical sharps and discarded swabs, while supporting a spill containment kit on its lower shelf and an active radiation monitoring station across its upper surface. Demarcated floor areas around the operating table were lined with non-slip, impervious Packexe decorator matting to catch fluid droplets and prevent radioactive contamination of structural theatre flooring.
Contamination monitoring performed with calibrated Radhound digital survey meters equipped with SS300 probes confirmed that radioactive contamination was generally restricted to low-level concentrations on disposable intraoperative consumables. Resected head and neck tissue specimens exhibited a mean surface dose rate of 1.1 microsieverts per hour and a mean calculated radioactivity of 170 kilobecquerels, whereas urology specimens demonstrated a mean surface dose rate of 0.9 microsieverts per hour and a mean activity of 33 kilobecquerels. Surgical swabs collected during head and neck cases showed contained radioactivity below 6 kilobecquerels. Conversely, urological swabs exhibited substantially higher activity concentrations due to rapid renal excretion of gallium-68 radiotracers into the urinary tract; peak swab readings reached 250 kilobecquerels at eighty-eight minutes post-injection and 160 kilobecquerels at 172 minutes, although most remained below 50 kilobecquerels. Only two minor personal contamination incidents occurred during the trial: one where approximately 10 kilobecquerels of gallium-68 contaminated an assistant surgeon's footwear following an accidental transection of a urinary catheter drainage tube, and another where 5 kilobecquerels of fluorine-18 adhered to a lead surgeon's surgical scrubs. Both staff members promptly removed the affected articles for radioactive decay storage, and subsequent skin surveys confirmed that no cutaneous contamination occurred.
The study also examined the secondary radiation exposure pathways affecting hospital staff located outside the primary operating environment, including post-anaesthetic recovery nurses, ward staff, and histopathology laboratory personnel. Personal electronic dosimeter measurements showed that recovery room nurses providing direct post-operative care received negligible radiation doses, averaging 0.7 microsieverts per patient during head and neck cases and 0.6 microsieverts per patient following urological cases. Investigators calculated that a post-anaesthetic nurse would need to care for more than 1,400 consecutive surgical patients annually before approaching a regulatory occupational exposure threshold of one millisievert. Resected tissue specimens were placed in formalin and held within a secure utility cabinet inside the operating theatre for twenty-four hours, allowing short-lived positron radioisotopes to undergo multiple half-life decays before transfer to pathology. By the time tissue arrived in the pathology laboratory, surface dose rates had diminished to background levels, measuring 0.4 nanosieverts per hour for fluorine-18 specimens and 0.00084 nanosieverts per hour for gallium-68 specimens, confirming that laboratory staff incurred no measurable occupational radiation hazard.
Translating the trial's dosimetric observations into annualized operational projections reveals that occupational exposure is heavily governed by surgical volume, procedural case mix, and institutional training compliance. In high-volume surgical centres performing an estimated twenty-four to forty-eight complex head and neck resections per year, primary operating surgeons could accumulate up to 1.8 millisieverts of whole-body exposure and 4.2 millisieverts of extremity exposure annually. Under prevailing United Kingdom radiation legislation, whole-body doses exceeding one millisievert designate an employee as someone who routinely works with ionizing radiation, potentially triggering formal occupational classification depending on institutional arrangements. In urology departments, where robotic procedures incur lower unit doses of 7.2 microsieverts, an assistant surgeon would need to participate in 138 procedures annually to reach a comparable one-millisievert threshold. The trial highlighted the critical role of continuous education and theatre-based cascade trainers, whose peer-led supervision reinforced distance discipline, personal monitoring routines, and contamination protocols during unexpected shift rotations, demonstrating that advanced specimen PET-CT imaging can be safely integrated into standard operating theatres without compromising staff safety.
Source
- Bioengineer.org (2026-11-16) - Original article: Specimen PET-CT Trial Confirms Surgical Safety Profiles for Operating Theatre Staff - https://bioengineer.org/specimen-pet-ct-trial-confirms-surgical-safety-profiles-for-operating-theatre-staff/